← All reports
PDF Excel ReqIF

Radio Chemistry Laboratory for a UK Nuclear Dockyard

Verification Plan (SVP) — ISO/IEC/IEEE 15289 — Plan | IEEE 29148 §6.6
Generated 2026-03-27 — UHT Journal / universalhex.org

82
Verification Entries
82
Verification Links
0
Orphans

Verification Requirements (VER)

RefRequirementMethodTags
VER-METHODS-001 Verify IFC-DEFS-001: Demonstrate flask docking operation with a standard 100L shielded transport flask containing a calibrated gamma source. Measure dose rate at operator positions during docking sequence using calibrated survey instruments. Pass criteria: no measured dose rate exceeds 7.5 microSv/h at any accessible surface; docking completes within 15 minutes from flask presentation to cell closure.
Rationale: Active commissioning test validates shielded flask docking mechanism and dose rate performance at the critical primary containment interface. Test with calibrated source confirms shielding effectiveness under representative loading conditions.
Test verification, hot-cell, session-226
VER-METHODS-002 Verify IFC-DEFS-002: Measure extract air flow rate at the cell-to-duct transition using a pitot tube traverse. Test isolation damper leak rate with the cell pressurised to +100 Pa. Pass criteria: flow rate within 500-1500 m3/h range; damper leakage less than 0.1% of rated flow; duct integrity confirmed by DOP aerosol challenge test.
Rationale: Pitot tube flow measurement and damper leak-rate testing confirm extract airflow maintains required negative pressure during normal and isolated conditions, validating the primary contamination containment mechanism at the cell-to-duct interface.
Test verification, hot-cell, session-226
VER-METHODS-003 Verify IFC-DEFS-003: Inject test signals at the in-cell instrument terminals simulating dose rates from 0.1 microSv/h to 10 Sv/h and a criticality alarm event. Measure signal arrival time at the central radiation protection system. Pass criteria: 4-20 mA alarm signals arrive within 100 ms; Modbus TCP data updates within 1 second; alarm state correctly displayed on central system.
Rationale: End-to-end signal injection testing validates the radiation monitoring interface from in-cell detector terminals through to the health physics display, covering the full dynamic range from background to criticality event levels including alarm trigger thresholds.
Test verification, hot-cell, session-226
VER-METHODS-004 Verify IFC-DEFS-004: Flow-test the active drain with clean water at the rated 5 L/min maximum flow rate for 30 minutes. Verify U-trap seal integrity by applying negative pressure to the drain outlet and confirming no air backflow through the seal. Pass criteria: sustained 5 L/min flow without overflow or blockage; U-trap holds seal under 200 Pa negative pressure.
Rationale: Flow testing with clean water at design maximum rate confirms drainage capacity and identifies potential blockage or backflow issues. U-trap seal integrity under negative pressure prevents contaminated air bypassing the ventilation containment barrier.
Test verification, hot-cell, session-226
VER-METHODS-005 Verify IFC-DEFS-005: Place a calibrated Cs-137 source of known activity at each manipulator slave arm position inside the cell. Measure dose rate at 30 cm from the manipulator wall penetration on the operator side using a calibrated ionisation chamber. Pass criteria: measured attenuation factor within 10% of calculated wall attenuation; no detectable contamination on the cold face of the manipulator boot seal after 100 operational cycles.
Rationale: Dose rate measurement at manipulator penetrations with a calibrated source inside the cell confirms radiation streaming through the labyrinth penetrations meets the shielding specification. This is the critical acceptance test for the shielding/penetration design.
Test verification, hot-cell, session-226
VER-METHODS-006 Verify SUB-REQS-006: Attempt to open both inner and outer transfer port doors simultaneously using manual and powered actuation. Confirm mechanical interlock prevents second door from opening while first door is open. Test all transfer ports. Pass criteria: no test achieves simultaneous opening of both doors under any actuation sequence.
Rationale: Interlock testing confirms the safety function preventing simultaneous opening of both transfer port doors, which would breach the primary containment barrier. Both manual and powered actuation modes must be tested as they use different actuation mechanisms.
Test verification, hot-cell, session-226
VER-METHODS-007 Verify SUB-REQS-008: Inject simulated criticality pulse signals into each criticality detection channel independently and simultaneously. Measure time from signal injection to alarm actuation. Confirm each channel operates independently. Pass criteria: alarm response within 500 ms on each channel; both channels produce alarms independently; alarm is annunciated at the central radiation protection system and local sounder.
Rationale: Criticality warning system response time testing validates the safety function from detection through alarm actuation. Independent and simultaneous channel testing confirms voting logic operates correctly under both single-channel and multi-channel exceedance conditions.
Test verification, hot-cell, session-226
VER-METHODS-008 Verify SUB-REQS-011: Simulate loss of cell depression by stopping the extract fan. Attempt to open each transfer port and the cell access door. Pass criteria: all transfer ports and cell access door remain locked when cell depression falls below 50 Pa; alarm is generated within 5 seconds of depression loss.
Rationale: Testing confirms the safety interlock preventing containment breach when cell depression is lost. This is a nuclear safety significant test — loss of depression combined with open transfer ports could allow contaminated air to escape the cell into occupied areas.
Test verification, hot-cell, session-226
VER-METHODS-009 Verify IFC-DEFS-007: Flow-test each active drain line with clean water at the maximum foreseeable flow rate. Confirm gravity flow from each laboratory source to the storage tank with no surcharging, measure flow gradient with survey level, and verify submerged termination by observing outlet below liquid level in tank at minimum operating level. Pass: all lines achieve gravity flow at design rate with gradient at or above 1:40.
Rationale: Flow testing at maximum foreseeable rate confirms gravity drainage capacity from each laboratory area to the collection tanks without surcharge. Adequate drainage prevents floor flooding with active liquids in areas where operators are present.
Test verification, aetp, session-227
VER-METHODS-010 Verify IFC-DEFS-008: Transfer a 2000L batch of clean water from storage tank to neutralisation vessel using the shielded centrifugal pump. Record flow totaliser reading at start and end, compare against calibrated volumetric measurement of receiving vessel. Operate remote isolation valves from control room. Pass: flow totalisation accurate to within 2%, remote valve operation confirmed, transfer rate within 10-50 L/min range.
Rationale: Full-scale batch transfer test validates pump capacity, pipeline flow resistance, and containment integrity of the shielded transfer line under representative operating conditions. Confirms the treatment plant can accept full tank batches within the operational time window.
Test verification, aetp, session-227
VER-METHODS-011 Verify IFC-DEFS-009: Operate evaporator with clean water feed and verify condensate delivery to IX columns at rated flow of 500 L/hr. Measure condensate temperature at IX inlet and confirm below 40C. Inject 100 mg/L NaCl solution into condensate and confirm conductivity alarm triggers within 30 seconds. Pass: flow rate achieved, temperature below limit, conductivity alarm functional at 50 mg/L threshold.
Rationale: Condensate flow and temperature testing validates the evaporator-to-IX interface performance. Condensate temperature and pH must be within IX resin tolerance to prevent resin degradation, which would compromise decontamination factors.
Test verification, aetp, session-227
VER-METHODS-012 Verify IFC-DEFS-010: Operate peristaltic pump with simulated concentrate at design viscosity through the shielded transfer line at 1-10 L/min. Execute remote flush cycle and confirm line clearance by visual inspection of flush water clarity. Verify shielding survey of transfer route shows dose rate below 7.5 microsieverts per hour at accessible positions. Pass: pump delivers at rated flow, flush achieves clear returns, shielding adequate.
Rationale: Peristaltic pump testing with simulated concentrate validates the shielded transfer line for viscous radioactive concentrate, including remote flushing capability that minimises operator dose during line clearance operations.
Test verification, aetp, session-227
VER-METHODS-013 Verify IFC-DEFS-011: Inject calibrated test signals into the monitoring station alpha and beta measurement channels simulating activities from background to 200% of trip level. Confirm 4-20 mA analogue signals track within 1% of injected values at the discharge control PLC input terminals. Confirm Modbus TCP data matches analogue within 0.5%. Simulate trip condition and confirm discharge valve closure within 5 seconds. Pass: analogue and digital channels agree, trip function operates within specified response time.
Rationale: Calibrated test signal injection validates the monitoring station measurement accuracy across the full range from background to 200% of discharge authorisation limits, ensuring reliable interlock activation before permit exceedance.
Test verification, aetp, session-227
VER-METHODS-014 Verify IFC-DEFS-012: Execute a simulated discharge batch on the control system and confirm the complete batch record (identifier, volume, radionuclide activities, timestamp, operator IDs) appears in LIMS within 60 seconds of batch completion. Verify data integrity by comparing all fields between source and destination. Pass: all required fields present in LIMS, data matches source, transfer within time limit.
Rationale: Simulated discharge batch validates the complete data chain from monitoring station through discharge control to LIMS record, confirming regulatory compliance traceability for every discharge event.
Demonstration verification, aetp, session-227
VER-METHODS-015 Verify SUB-REQS-016: Perform a SIL 2 proof test of the discharge safety function by injecting calibrated high-activity test signals into all monitoring channels simultaneously. Measure time from signal injection to confirmed valve closed position using high-speed data logger. Repeat for each individual channel. Verify spurious trip rate is below 1 per year by analysis of 1000-hour soak test data. Pass: valve closure within 5 seconds on all test scenarios, SIL 2 PFDavg demonstrated by analysis.
Rationale: SIL 2 proof test validates the discharge safety function by demonstrating correct response to high-activity signals, confirming the safety interlock prevents environmental discharge above permitted activity levels.
Test verification, aetp, session-227
VER-METHODS-016 Verify IFC-DEFS-013: Measure duct velocity at the HEPA filter bank inlet using a pitot tube traverse across the 300mm duct cross-section at design extract flow rate. Confirm velocity does not exceed 10 m/s. Operate each isolation damper to full closed and full open positions and verify full stroke in less than 30 seconds. Pass criteria: duct velocity 10 m/s or less, damper full stroke within 30 seconds, no air leakage past closed damper exceeding 0.5% of rated flow.
Rationale: Pitot tube velocity measurement at the HEPA bank inlet confirms uniform airflow distribution across the filter face, which is essential for achieving the specified filtration efficiency and preventing premature filter breakthrough.
Test verification, ventilation, session-228
VER-METHODS-017 Verify IFC-DEFS-014: Inject a step change in fume cupboard sash position (full open to half-open) and record the Pressure Cascade Control System response. Verify the 4-20 mA control signal to the extract fan VSD changes within 1 second and that the laboratory pressure differential returns to within 10% of set-point within 5 seconds. Pass criteria: control signal update within 1 second, pressure recovery within 5 seconds, no pressure reversal during transient.
Rationale: Dynamic response testing of the pressure cascade control system validates automatic compensation for sash position changes, ensuring the fume cupboard face velocity and room pressure cascade are maintained during normal laboratory operations.
Test verification, ventilation, session-228
VER-METHODS-018 Verify IFC-DEFS-015: Configure a test data transmission between the Stack Monitoring Instrumentation and LIMS. Inject calibrated alpha, beta, and iodine test signals into the monitoring instruments. Verify Modbus TCP data records arrive in LIMS at 15-minute intervals containing timestamp, cumulative activity, and instrument status fields. Pass criteria: all 3 radionuclide categories received, timestamp accuracy within 2 seconds of NTP reference, instrument status correctly reflects calibration mode.
Rationale: End-to-end data transmission testing from stack monitors to LIMS validates automated discharge reporting, eliminating the risk of manual transcription errors in environmental permit compliance data.
Test verification, ventilation, session-228
VER-METHODS-019 Verify IFC-DEFS-016: Measure supply air velocity at each ceiling diffuser face using a calibrated thermal anemometer with the extract system operating at design flow rate and fume cupboard sashes at normal working position. Pass criteria: no diffuser face velocity exceeds 0.3 m/s, air distribution pattern does not create cross-drafts at fume cupboard openings exceeding 0.1 m/s.
Rationale: Supply air velocity measurement at each diffuser confirms balanced air distribution providing the clean air supply that drives the pressure cascade from clean to active areas. Uniform distribution prevents stagnant zones where contamination could accumulate.
Test verification, ventilation, session-228
VER-METHODS-020 Verify IFC-DEFS-017: Inject calibrated test signals at 7.5 microsievert/h and 2 mSv/h thresholds into each area gamma detector channel. Measure time from signal injection to alarm relay actuation at the central monitoring panel. Pass criteria: alarm actuation within 2 seconds for all channels; analogue alarm channel actuates independently when Modbus TCP link is disconnected.
Rationale: Tests at both alarm thresholds (7.5 μSv/h and 2 mSv/h) verify the full safety-critical signal path from detector through analogue channel to the centralised display and alarm system, confirming end-to-end functionality under the dual-path architecture specified in IFC-DEFS-017.
Test verification, radpro, session-229
VER-METHODS-021 Verify IFC-DEFS-018: Set test worker cumulative dose to 9.9 mSv in the dosimetry database and confirm controlled area entry is permitted. Increase to 10.1 mSv and confirm entry is denied at the turnstile within 500 ms of badge presentation. Verify dose query response time under simulated peak load of 20 simultaneous queries.
Rationale: Testing at 9.9 mSv (just below the 10 mSv threshold) and then at 10.0 mSv verifies the access control boundary condition, ensuring the system correctly prevents entry at the dose constraint and permits entry just below it. This tests the critical integration between Personal Dosimetry Management and Radiological Access Control per IFC-DEFS-018.
Test verification, radpro, session-229
VER-METHODS-022 Verify IFC-DEFS-019: De-energise the general area monitoring system and the facility IT network. Inject a simulated criticality signal into the criticality warning system detectors. Confirm evacuation sounders and beacons activate within 500 ms. Introduce open-circuit and short-circuit faults on alarm wiring and confirm fault detection within 60 seconds.
Rationale: De-energising the general monitoring and IT systems before testing verifies the criticality warning system's independence claim per IFC-DEFS-019. This test demonstrates that the hardwired relay path operates without any dependence on the systems it must be independent of — the most stringent test of the independence requirement.
Test verification, radpro, criticality, session-229
VER-METHODS-023 Verify IFC-DEFS-020: Generate a test alarm event on the central monitoring system. Confirm the alarm record appears in the LIMS within 60 seconds with correct timestamp, source identifier, measurement value, and alarm status. Verify 15-minute routine data transfer includes all active monitoring channels.
Rationale: Verifies the alarm event transmission path from the centralised monitoring system to LIMS per IFC-DEFS-020, confirming that alarm events are received within the 60-second requirement and contain the complete data fields needed for regulatory reporting.
Test verification, radpro, session-229
VER-METHODS-024 Verify IFC-DEFS-021: Inject a simulated alpha activity signal exceeding 1 DAC into a continuous air monitor in a test zone. Confirm that the hardwired trip signal is received by the Pressure Cascade Control System and that extract flow rate increases by at least 50% within 30 seconds. Verify trip signal operates with digital network disconnected.
Rationale: Verifies the hardwired trip signal path from continuous air monitor to ventilation system per IFC-DEFS-021, testing the full chain from contamination detection through automatic extract flow increase, confirming the 30-second response time at the 1 DAC threshold.
Test verification, radpro, ventilation, session-229
VER-METHODS-025 Verify SUB-REQS-035: Using a Cf-252 neutron source and Co-60 gamma source at calibrated positions, demonstrate that the criticality warning system 2-out-of-3 voting logic actuates the evacuation alarm within 500 ms of both gamma and neutron thresholds being exceeded simultaneously. Repeat for each detector combination to verify all 2-of-3 permutations.
Rationale: Using Cf-252 (neutron) and Co-60 (gamma) calibration sources at known positions verifies the detector response, coincidence voting logic, and alarm actuation time specified in SUB-REQS-035. Testing in the actual monitored zone geometry confirms performance under real-world conditions including reflections and shielding effects.
Test verification, radpro, criticality, session-229
VER-METHODS-026 Verify SUB-REQS-038: Using NIST-traceable alpha (Am-241) and beta (Sr-90/Y-90) calibration sources at activities corresponding to 0.4 Bq/cm2 alpha and 4 Bq/cm2 beta, demonstrate that all hand and clothing monitors detect and alarm within 30 seconds. Verify exit gate interlock prevents release when contamination is detected.
Rationale: NIST-traceable calibration sources provide known reference values for verifying the detection sensitivity thresholds specified in SUB-REQS-038. Am-241 and Sr-90/Y-90 sources test alpha and beta detection pathways respectively, representing the two contamination types the monitors must detect at the specified clearance levels.
Test verification, radpro, contamination, session-229
VER-METHODS-027 Verify IFC-DEFS-022: Transfer 10 representative waste items from the Sorting Facility to the NDA System via shielded trolley. Confirm each item's digital identification tag is read successfully by the NDA barcode scanner on first attempt. Pass criteria: 100% successful tag reads with correct item identification.
Rationale: Testing with 10 representative waste items validates the complete sorting-to-assay transfer chain per IFC-DEFS-022, including shielded trolley transfer, barcode scanning, and item identification. Representative items cover the range of geometries and activity levels expected in routine operations.
Test verification, solid-waste, session-230
VER-METHODS-028 Verify IFC-DEFS-023: Perform an NDA measurement on a calibration drum and verify that the complete assay result set (package ID, date, radionuclide inventory, uncertainties, total activity, fissile mass) appears in the Waste Tracking System within 60 seconds. Repeat for 5 consecutive measurements. Pass criteria: all 5 data transfers complete within 60 seconds with zero field errors.
Rationale: Using a calibration drum with known radionuclide content provides a reference for verifying the complete NDA-to-tracking data transfer per IFC-DEFS-023, including all required data fields, structured format, and 60-second transfer time. Calibration drums provide traceable reference values against which data completeness and accuracy can be independently verified.
Test verification, solid-waste, session-230
VER-METHODS-029 Verify IFC-DEFS-024: Transfer an empty HHISO container from the compaction area to the waste store via forklift through the loading bay. Measure clearances at all points. Confirm the container identifier and storage location are recorded in the Waste Tracking System upon placement. Pass criteria: minimum 300 mm clearance each side, tracking record created within 5 minutes of placement.
Rationale: Testing with an actual HHISO container validates the physical transfer path (loading bay clearance, forklift access) and the digital recording chain (container ID, storage location) per IFC-DEFS-024. Using an empty container provides a safe test while exercising the full physical and data transfer interface.
Inspection verification, solid-waste, session-230
VER-METHODS-030 Verify IFC-DEFS-025: Perform a simulated ILW drum transfer from the conditioning plant to the ILW storage area using the overhead crane. Verify drum identifier, weight, and surface dose rate are recorded in the Waste Tracking System before crane release at the storage location. Pass criteria: all three data fields populated before placement, crane load verified within 2% of weigh-scale reading.
Rationale: Simulated ILW drum transfer exercises the crane handling path and complete data recording chain per IFC-DEFS-025, verifying that weight, dose rate, and conditioning date are captured before storage placement. A simulation rather than active drum is used because the test must verify the interface mechanics before active commissioning.
Test verification, solid-waste, session-230
VER-METHODS-031 Verify IFC-DEFS-026: Submit a test set of 5 radiochemical analysis results from the LIMS for a known waste package identifier. Verify results appear in the Waste Tracking System linked to the correct package. Submit a sample request from the Waste Tracking System and verify it appears in the LIMS work queue. Pass criteria: bidirectional data exchange with zero linkage errors.
Rationale: Testing with 5 known waste packages verifies the bidirectional LIMS-Waste Tracking data exchange per IFC-DEFS-026, confirming that radiochemical analysis results flow from LIMS to Waste Tracking and that sample-to-package traceability is maintained. A known dataset enables independent verification of data completeness and correctness.
Demonstration verification, solid-waste, session-230
VER-METHODS-032 Verify SUB-REQS-031: Position NIST-traceable Cs-137 calibration sources at each detector location. Confirm each ionisation chamber reads within 15% of the reference dose rate across five decades from 0.1 microsievert/h to 10 sievert/h. Verify all occupied zones, hot cell positions, and waste handling areas have at least one functioning detector. Pass criteria: all detectors operational and within calibration tolerance with no coverage gaps.
Rationale: NIST-traceable source calibration check confirms each detector reads within the specified accuracy band, validating the area monitoring network's ability to provide reliable dose rate measurements for worker protection decisions.
Test verification, radpro, session-231
VER-METHODS-033 Verify SUB-REQS-032: Ramp a calibrated gamma source at one detector location through the 7.5 microsievert/h investigation threshold then through the 2 millisievert/h evacuation threshold. Measure time from threshold crossing to alarm actuation. Pass criteria: local alarm actuates within 5 seconds at both levels with correct alarm type annunciated.
Rationale: Ramp testing through alarm thresholds validates the full alarm chain from detector through voting logic to audible/visual alarm activation, confirming response times and correct threshold settings for both investigation and evacuation levels.
Test verification, radpro, session-231
VER-METHODS-034 Verify SUB-REQS-036: Isolate the criticality warning system UPS from mains supply and confirm all detectors, voting logic, and alarm circuits remain operational for 8 hours on battery. Separately disconnect all cables between the Area Gamma Dose Rate Monitoring Network and the CWS and confirm CWS continues to function. Verify self-test cycle completes within 24 hours and correctly reports a simulated detector fault. Pass criteria: full independence demonstrated, 8-hour battery endurance achieved, self-test detects injected fault.
Rationale: UPS endurance testing confirms the criticality warning system maintains full functionality during mains power loss. This is a safety-critical system that must operate continuously regardless of site power status.
Test verification, radpro, session-231
VER-METHODS-035 Verify SUB-REQS-037: Issue an electronic personal dosemeter to a test worker. Expose the dosemeter to a calibrated Cs-137 field at known dose rates of 1 microsievert/h, 100 microsievert/h, and 1 sievert/h. Confirm real-time display reads within 10% of reference. Set cumulative dose to trigger 0.5 mSv and 2 mSv alarm thresholds and confirm audible alarm activates at each level. Exit the controlled area and confirm automatic upload of dose record to the central system. Pass criteria: display accuracy within 10%, both alarms trigger, data uploads within 60 seconds of exit.
Rationale: EPD calibration verification confirms electronic personal dosemeters read accurately across the dose rate range encountered in the laboratory, from background levels in inactive areas to elevated rates near active waste stores.
Test verification, radpro, session-231
VER-METHODS-036 Verify SUB-REQS-041: Present 20 representative waste items covering all expected categories (VLLW, LLW compactable, LLW non-compactable, ILW) to the Waste Sorting and Segregation Facility. Confirm all items are categorised within 24 hours of receipt. Verify categorisation against independent radiometric characterisation. Pass criteria: all items categorised within 24 hours, correct waste stream assignment for all items, contamination and dose rate measurements recorded for each item.
Rationale: Waste sorting accuracy testing with representative waste categories validates the classification system's ability to correctly categorise waste for the appropriate disposal route, preventing regulatory non-compliance from miscategorised packages.
Test verification, solid-waste, session-231
VER-METHODS-037 Verify SUB-REQS-042: Perform segmented gamma scanning of 5 NIST-traceable calibration drums containing known activities of Co-60, Cs-137, and Am-241 in both uniform and non-uniform distributions. Confirm measurement uncertainty is less than 20% at 2-sigma for each radionuclide contributing more than 1% of total activity. Verify 12 axial segments are acquired with full drum rotation. Pass criteria: all reported uncertainties within specification for all calibration drums.
Rationale: Segmented gamma scanning verification with NIST-traceable calibration drums confirms measurement accuracy for waste characterisation across the activity range and radionuclide mix expected from radiochemistry operations.
Test verification, solid-waste, session-231
VER-METHODS-038 Verify SUB-REQS-033: Inject calibrated alpha (Am-241) and beta (Sr-90) aerosol sources into each active laboratory zone continuous air monitor. Confirm minimum detectable concentration of 0.02 DAC for alpha and 0.1 DAC for beta. Verify measurement cycle completes within 10 minutes. Verify radon background compensation by measuring with and without radon-laden air. Pass criteria: MDC values achieved, cycle time within spec, radon compensation reduces false-positive rate.
Rationale: Aerosol injection testing validates continuous air monitor sensitivity for alpha and beta contamination detection, confirming the system can detect airborne contamination events that would trigger area evacuation and investigation.
Test verification, radpro, session-231
VER-METHODS-039 Verify IFC-DEFS-027: Conduct integration test by transmitting 100 sample registration messages from the Logging Station to LIMS over a 60-minute period. Measure end-to-end message latency for each. Simulate LIMS server unavailability for 30 seconds and confirm automatic retry delivers all pending messages. Pass criteria: all 100 messages received with correct content, latency below 2 seconds for 99 percent of messages, zero lost messages after retry.
Rationale: Integration testing of the Sample Logging Station to LIMS data path verifies end-to-end message integrity, latency compliance, and error recovery. The 100-message test provides statistical confidence in the 2-second latency requirement across varying network loads.
Test verification, sample-prep, session-232
VER-METHODS-040 Verify IFC-DEFS-028: Measure extract flow rate at the duct connection using a calibrated vane anemometer or pitot tube traverse at commissioning and annually thereafter. Verify flow rate at 0.35 m3/s minimum with fume cupboard sash at maximum working height. Inspect duct integrity and damper operation. Pass criteria: flow rate at or above 0.35 m3/s, no visible duct damage or leakage, damper moves freely through full range.
Rationale: Direct airflow measurement at the duct connection validates the extract system's ability to maintain containment velocity at the fume cupboard. Calibrated vane anemometry provides UKAS-traceable measurement of the 0.35 m³/s flow requirement.
Test verification, sample-prep, session-232
VER-METHODS-041 Verify IFC-DEFS-029: Inspect drain installation for correct material specification (316L SS), trapped seal presence, and sampling point accessibility. Perform leak test at 0.5 bar positive pressure. Pass criteria: no visible defects, seal holds without drain-through for 24 hours, sampling valve operable, leak test holds for 30 minutes.
Rationale: Physical inspection verifies material specification (316L SS), trap seal integrity, and sampling point accessibility — attributes that cannot be verified by functional test alone. Material certificates and dimensional records form part of the commissioning quality dossier.
Inspection verification, sample-prep, session-232
VER-METHODS-042 Verify IFC-DEFS-030: Conduct transfer test using sealed dummy containers of 250ml volume containing tracer solution. Measure transfer time for 10 consecutive transfers. Swab outer container surfaces post-transfer for contamination. Pass criteria: all transfers complete within 3 minutes, all outer surfaces below 4 Bq/cm2 beta-gamma and 0.4 Bq/cm2 alpha.
Rationale: Transfer testing with sealed dummy containers validates the pneumatic/carrier system performance without radiological hazard. The 250mL volume and containment seal check verify the interface against the maximum design envelope and double-seal integrity requirement.
Test verification, sample-prep, session-232
VER-METHODS-043 Verify IFC-DEFS-031: Measure extract flow rate at duct connection with oven at operating temperature (900 deg C furnace). Verify negative pressure differential between oven enclosure and laboratory using manometer. Inspect HEPA filter rating label and installation seal. Pass criteria: flow rate at or above 0.1 m3/s, negative pressure of at least minus 10 Pa relative to lab, HEPA filter correctly rated and sealed.
Rationale: Measuring extract flow with the oven at operating temperature validates thermal conditions do not degrade ductwork performance or HEPA filter integrity. This is a functional test that cannot be replaced by design analysis because thermal expansion and filter pressure drop are installation-specific.
Test verification, sample-prep, session-232
VER-METHODS-044 Verify IFC-DEFS-032: Connect balance to LIMS test instance. Perform 20 weighings of calibration masses. Confirm all 20 records appear in LIMS with correct mass values, units, timestamps, balance ID, and temperature. Introduce a deliberate communication break and confirm data is buffered and retransmitted. Pass criteria: 100 percent data capture accuracy, all fields populated, no data loss after reconnection.
Rationale: Serial communication testing with calibration masses of known value verifies both the data link integrity and the accuracy of transmitted measurement records. The 20-weighing sample size provides confidence in reliability and the inclusion of stability and temperature fields validates completeness of the data record.
Test verification, sample-prep, session-232
VER-METHODS-045 Verify SUB-REQS-050: Position calibration sources (Cs-137, Co-60) at 100mm from the reception bay dose rate instrument. Confirm instrument reads within plus or minus 20 percent of known dose rate across the range 0.1 uSv/h to 10 mSv/h. Time 10 consecutive dose rate measurements and confirm all complete within 60 seconds. Pass criteria: all readings within calibration tolerance, all measurements within 60-second limit.
Rationale: Using calibration sources of known dose rate (Cs-137, Co-60) at the specified 100mm geometry validates the instrument's measurement accuracy and range compliance. This directly tests the requirement's stated performance envelope at the actual installation position.
Test verification, sample-prep, session-232
VER-METHODS-046 Verify SUB-REQS-051: During a simulated campaign period, process 80 dummy samples through the full preparation workflow (receipt, logging, storage, preparation, aliquoting) over a 5-day period. Record time from receipt to delivery of prepared aliquots for each sample. Pass criteria: all 80 samples processed, 95 percent within 48-hour turnaround.
Rationale: Processing 80 dummy samples through the full preparation workflow during a simulated campaign demonstrates throughput capacity under realistic conditions, including staff shift patterns and equipment turnaround. Demonstration is the appropriate verification method because throughput is an emergent property of the whole preparation system, not testable at component level.
Demonstration verification, sample-prep, session-232
VER-METHODS-047 Verify SUB-REQS-059: Present check sources of known activity to the contamination monitoring instruments. Confirm detection of 0.4 Bq/cm2 alpha using Am-241 source and 4 Bq/cm2 beta-gamma using Cl-36 source with instrument alarm actuation. Pass criteria: audible and visual alarm triggers at or below the specified contamination limits, instrument responds within 10 seconds.
Rationale: Check sources of known activity verify the contamination monitoring instruments' detection sensitivity against the 0.4 Bq/cm² alpha and 4 Bq/cm² beta-gamma thresholds. This is a direct functional test of the measurement capability that underpins the waste segregation decision point.
Test verification, sample-prep, session-232
VER-METHODS-048 Verify IFC-DEFS-033: Measure extract airflow at each fume cupboard branch duct using a calibrated pitot traverse at commissioning and annually thereafter. Pass criterion: 0.6 m3/s plus or minus 10% per cupboard with sash fully open. Verify HEPA filter integrity by in-situ DOP aerosol challenge test to BS EN 1822, pass criterion: penetration less than 0.003%.
Rationale: Pitot tube measurement at each branch duct verifies that the extract system delivers the required 0.6 m³/s per cupboard when all cupboards are operating simultaneously. This tests the balanced airflow distribution which is critical for containment and cannot be verified by single-point measurement.
Test verification, radiochem-sep, session-233
VER-METHODS-049 Verify IFC-DEFS-034: Pressure-test the borosilicate glass drain pipework at 1.5 bar gauge for 30 minutes at commissioning. Pass criterion: no visible leakage or pressure decay greater than 0.05 bar. Verify gradient by surveyed level measurement at pipe entry and exit points. Verify material compatibility by exposing test coupons to 8M HNO3 at 80 degrees C for 168 hours with mass loss less than 0.1 mg per square centimetre.
Rationale: Hydrostatic pressure testing at 1.5 bar for 30 minutes verifies the integrity of borosilicate glass drain pipework joints before commissioning with active liquors. This is the standard pipeline integrity test for acid-resistant glass systems and detects joint defects that would cause leakage of radioactive effluent.
Test verification, radiochem-sep, session-233
VER-METHODS-050 Verify IFC-DEFS-035: Conduct a blind sample transfer exercise with 20 uniquely barcoded vials transferred from the separations laboratory to each analytical instrument. Pass criterion: 100% correct identification at receiving end with complete LIMS records for all 20 transfers.
Rationale: Blind sample transfer with barcoded vials verifies end-to-end chain of custody between the extraction chromatography station and receiving analytical instruments. The 20-vial test size provides statistical confidence in barcode readability and LIMS tracking accuracy across all four downstream instrument destinations.
Test verification, radiochem-sep, session-233
VER-METHODS-051 Verify IFC-DEFS-036: Test the interlocked pass-through hatch by attempting simultaneous opening of both doors. Pass criterion: interlock prevents opening of second door in all 10 test cycles. Verify pressure cascade maintenance by measuring differential pressure across the hatch during door operation, pass criterion: pressure does not reverse during any phase of transfer.
Rationale: Testing the interlocked pass-through hatch by attempting simultaneous opening verifies the safety interlock's ability to prevent breach of the contamination boundary. This is a direct safety function test that must be demonstrated to pass before the facility receives its commissioning certificate.
Test verification, radiochem-sep, session-233
VER-METHODS-052 Verify IFC-DEFS-037: Dispense 10 tracer aliquots in sequence, verifying that LIMS records each dispensing event with correct tracer identity, operator, volume, and timestamp. Pass criterion: all 10 events recorded within 30 seconds of dispensing and all fields match physical records. Verify expired-tracer lockout by attempting to scan an expired barcode.
Rationale: Sequential tracer dispensing with LIMS verification tests the real-time inventory tracking system under typical operational conditions. Verifying each dispensing event within the 30-second latency requirement validates that the system can keep pace with batch preparation workflows.
Test verification, radiochem-sep, session-233
VER-METHODS-053 Verify IFC-DEFS-038: Analyse 5 reference spectra of known activity and verify that results transferred to LIMS match the analysis software output to within rounding precision. Verify checksum by intentionally corrupting one transfer file and confirming rejection. Pass criterion: 5/5 correct transfers and 1/1 corruption detected.
Rationale: Analysing reference spectra of known activity provides a direct test of the automated data bridge accuracy, including nuclide identification, activity quantification, and checksum verification. The 5-spectrum test covers the key radionuclides encountered in dockyard environmental and waste samples.
Test verification, gamma-spec, session-233
VER-METHODS-054 Verify IFC-DEFS-039: Load 10 barcoded source discs of 25mm diameter into the sample changer carousel. Verify that the barcode reader correctly identifies each disc and that LIMS records match the expected source identifier, nuclide group, and electrodeposition date. Confirm mechanical compatibility by verifying each disc seats correctly in the carousel position without manual intervention.
Rationale: Sample changer automation testing confirms barcode-driven sample identification and positioning accuracy, preventing cross-contamination and sample mix-up errors in alpha spectrometry measurements.
Test verification, alpha-spec, session-234
VER-METHODS-055 Verify IFC-DEFS-040: Analyse 5 reference spectra of known activity and verify that the automated data bridge transfers nuclide identities, activity concentrations, uncertainties, yield values, and spectrum file references to the LIMS test instance. Confirm checksum verification by corrupting one transfer and verifying the system detects and rejects it.
Rationale: Data bridge validation with reference spectra confirms accurate transfer of nuclide identities, activities, and uncertainties from the gamma spectrometry analyser to LIMS, maintaining measurement traceability through the automated data chain.
Test verification, alpha-spec, session-234
VER-METHODS-056 Verify SUB-REQS-076: Measure the FWHM energy resolution of each of the 8 PIPS detectors using a certified Am-241 point source (5.486 MeV alpha) at the standard 15mm source-to-detector distance. Record peak width at half maximum in keV. Pass criterion: all 8 detectors achieve 18 keV FWHM or better.
Rationale: FWHM energy resolution measurement with a certified Am-241 source validates each PIPS detector's spectroscopic performance, confirming ability to resolve adjacent alpha peaks from actinide isotopes in mixed-radionuclide samples.
Test verification, alpha-spec, session-234
VER-METHODS-057 Verify IFC-DEFS-041: Perform controlled transfer of 10 enriched tritium samples with known enrichment factors. Confirm each sample arrives with data sheet, batch number, and recovery factor. Verify enrichment factor accuracy by comparing back-calculated activity against a reference standard spiked prior to enrichment. Pass criteria: all 10 samples have complete documentation and back-calculated activities within 10% of reference value.
Rationale: Controlled transfer testing validates the enriched tritium sample interface, confirming quantitative sample transfer with associated metadata for traceability through the tritium analysis chain.
Test verification, lsc, session-235
VER-METHODS-058 Verify IFC-DEFS-042: Export raw spectral data from counter for 5 samples spanning the quench range (SQP(E) 200-800). Import into analysis software and verify: all 1024 channels present, SQP(E) values match counter display within 1 unit, counting times match, and sample IDs are correctly mapped. Pass criteria: zero data loss across all 5 test samples.
Rationale: Multi-quench-level spectral data import validates the analysis software's ability to handle the full range of sample quench conditions encountered in radiochemistry, ensuring accurate MDA and activity calculations across all sample types.
Test verification, lsc, session-235
VER-METHODS-059 Verify IFC-DEFS-043: Process 20 routine analytical results through the software-to-LIMS interface. Compare LIMS-stored values against software output for activity, uncertainty, MDA, and QC flags. Inject 2 results with deliberate QC failures and verify LIMS correctly flags them. Pass criteria: zero transcription discrepancies across all 20 results, both QC failures correctly flagged.
Rationale: 20-sample interface comparison validates automated result transfer accuracy between LSC analysis software and LIMS, confirming no data corruption or rounding errors in the automated data chain for regulatory reporting.
Test verification, lsc, session-235
VER-METHODS-060 Verify IFC-DEFS-044: Prepare 6 test fractions at acid normalities of 0.5M, 1.0M, 1.5M, 2.0M, 2.5M, and 3.0M HCl. Mix each with Ultima Gold LLT cocktail at standard ratio. Visually inspect for phase separation after 1 hour and after 24 hours. Pass criteria: no phase separation at or below 2.0M, phase separation observed above 2.0M confirming the limit is correctly specified.
Rationale: Acid normality compatibility testing confirms the LSC cocktail produces stable, clear mixtures across the range of acid strengths from radiochemical separations, preventing phase separation that would invalidate counting results.
Test verification, lsc, session-235
VER-METHODS-061 Verify SUB-REQS-081: Measure background count rate in the tritium window (0-18.6 keV) using sealed low-potassium glass blank vials counted for 1000 minutes on each counter. Repeat on 3 consecutive days. Pass criteria: mean background below 1.0 CPM with relative standard deviation below 15% across all measurements.
Rationale: 1000-minute background counting validates the ultra-low-background performance that defines this counter's capability for environmental-level tritium measurements. Background count rate directly determines the minimum detectable activity.
Test verification, lsc, session-235
VER-METHODS-062 Verify SUB-REQS-083: Process 6 environmental water samples spiked with known tritium activity (approximately 5 Bq/L) through the full enrichment cycle. Measure deuterium spike recovery and calculate enrichment factor. Back-calculate original activity and compare against known spike. Pass criteria: enrichment factor >= 20, deuterium recovery 90-110%, back-calculated activity within 15% of known spike.
Rationale: Spiked environmental water sample testing validates the full enrichment-and-counting chain from sample receipt through to reported activity, confirming the 1 Bq/L MDA is achieved end-to-end, not just instrument-level.
Test verification, lsc, session-235
VER-METHODS-063 Verify IFC-DEFS-045: Prepare 5 test actinide fractions from extraction chromatography with known TDS levels (0.05%, 0.1%, 0.15%, 0.2%, 0.5% m/v). Analyse each on ICP-MS and compare signal recovery against a clean 2% HNO3 calibration standard. Pass criteria: fractions at or below 0.1% TDS show signal recovery within 5% of standard; fractions above 0.1% show measurable suppression confirming the limit specification.
Rationale: TDS tolerance testing validates the clean room sample preparation enclosure delivers samples within the ICP-MS matrix tolerance, preventing signal suppression and transport efficiency losses that degrade measurement accuracy.
Test verification, icp-ms, session-235
VER-METHODS-064 Verify IFC-DEFS-046: Acquire 10 replicate analyses of IRMM-184 natural uranium standard at 1 ng/mL. Export time-resolved data via software interface. Verify 100 ms integration periods are present, internal standard count rates are recorded, and tuning parameters are included. Apply particulate exclusion algorithm and confirm isotope ratio precision matches instrument specification. Pass criteria: all data fields present, U-235/U-238 ratio within 0.3% of certified value.
Rationale: Replicate analysis of IRMM-184 uranium standard validates ICP-MS precision and the time-resolved data interface, confirming isotope ratio measurement capability meets safeguards-grade accuracy requirements.
Test verification, icp-ms, session-235
VER-METHODS-065 Verify IFC-DEFS-047: Process 10 IDMS results through the software-to-LIMS interface. Verify all fields (isotope ratios, IDMS concentrations, individual uncertainty components, expanded uncertainty) are correctly transferred. Retrieve records after 30 days and confirm data integrity is maintained. Pass criteria: zero discrepancies between software output and LIMS-stored values across all fields.
Rationale: IDMS result transfer validation confirms isotope ratio calculations, concentrations, and uncertainties are correctly passed through the software-to-LIMS interface without data loss, maintaining traceability for nuclear material accountancy.
Inspection verification, icp-ms, session-235
VER-METHODS-066 Verify IFC-DEFS-048: Inject 500 test result records from 5 simulated instruments simultaneously into the Instrument Data Acquisition Gateway. Confirm all 500 records appear in the database within 10 minutes with correct field mapping. Interrupt the network connection mid-batch and confirm no partial records exist after reconnection and synchronisation. Pass criteria: zero data loss, zero partial records, throughput >= 50 records/minute sustained.
Rationale: Concurrent multi-instrument data injection stress-tests the gateway's throughput capacity and confirms data integrity under peak load conditions representative of simultaneous analytical runs across the laboratory.
Test verification, lims, session-236
VER-METHODS-067 Verify IFC-DEFS-049: Attempt to submit an analytical result for a sample identifier that does not exist in the tracking module. Confirm the result is rejected with an appropriate error. Then register the sample and resubmit — confirm acceptance. Submit a priority sample and confirm the priority flag propagates to the results review queue. Pass criteria: 100% rejection of orphaned results, correct priority propagation.
Rationale: Negative testing confirms the sample tracking module rejects orphan results, preventing analytical data from being recorded without a valid chain-of-custody record. This is essential for forensic-quality sample traceability.
Test verification, lims, session-236
VER-METHODS-068 Verify IFC-DEFS-050: Create 20 test results spanning liquid effluent, gaseous, and solid waste pathways. Authorise all through the two-stage workflow. Confirm the regulatory module aggregates by pathway correctly. Attempt to modify a source result via the regulatory module interface — confirm the attempt is rejected. Submit a result that has only single-stage review — confirm it is excluded from regulatory aggregation. Pass criteria: correct pathway aggregation, zero unauthorised modifications, zero inclusion of unreviewed results.
Rationale: Multi-pathway result transfer testing validates the authorisation workflow and regulatory module integration across all waste discharge categories, confirming approved results correctly contribute to cumulative discharge totals.
Test verification, lims, session-236
VER-METHODS-069 Verify IFC-DEFS-051: Submit a sequence of 20 calibration check results for a test instrument, with results 18 and 19 deliberately exceeding the 3-sigma action limit. Confirm the control chart updates automatically, the action limit trigger generates a notification, and the instrument is locked out from new analytical runs. Clear the lockout and confirm the instrument resumes accepting runs. Pass criteria: real-time chart update, correct threshold detection, instrument lockout enforced, lockout clearance functional.
Rationale: Deliberate out-of-tolerance calibration check testing validates the QA module's automated flagging and instrument suspension capability, preventing results from miscalibrated instruments entering the quality system.
Test verification, lims, session-236
VER-METHODS-070 Verify IFC-DEFS-052: Log into a test workstation configured as an active area terminal. Leave the session idle for 14 minutes — confirm the session remains active. Wait one additional minute — confirm automatic logout. Repeat for a non-active area terminal at 29 and 31 minutes. Capture network traffic during a session and confirm TLS 1.2+ encryption with mutual certificate authentication. Attempt login from a device without a valid client certificate — confirm rejection. Pass criteria: correct timeout enforcement, TLS 1.2+ confirmed, unauthenticated device rejected.
Rationale: Session timeout and authentication testing validates cybersecurity controls at the operator workstation level, confirming unattended terminals are secured and all access events are logged for audit trail requirements.
Test verification, lims, session-236
VER-METHODS-071 Verify SUB-REQS-001: Perform gamma dose rate survey of all accessible external surfaces of the biological shielding structure using calibrated ionisation chamber instruments (range 0.1-1000 µSv/h) with sealed gamma sources of known activity (Co-60 and Cs-137 up to 1E14 Bq equivalent) positioned at representative cell interior locations. Pass criteria: no measured dose rate exceeds 7.5 µSv/h at any accessible surface. Survey to be conducted at commissioning and after any structural modification.
Rationale: External surface dose rate survey is the primary acceptance test for the biological shielding structure, confirming radiation exposure to operators in occupied areas adjacent to the hot cell meets the design dose rate limit.
Test verification, hot-cell, session-237
VER-METHODS-072 Verify SUB-REQS-004: With the in-cell ventilation extract system operating at design duty and all cell penetrations sealed, measure the differential pressure between the hot cell interior and the adjacent operator area using calibrated differential pressure transmitters at three locations. Pass criteria: all readings show at least 50 Pa negative differential sustained for a minimum of 60 minutes. Test with cell access door closed and all transfer ports in normal configuration.
Rationale: Differential pressure measurement validates the primary containment mechanism — negative pressure prevents contaminated air escaping the cell. This is the fundamental safety demonstration for the hot cell containment system.
Test verification, ventilation, session-237
VER-METHODS-073 Verify SUB-REQS-005: Perform in-situ aerosol challenge test on each HEPA filter stage using DOP or PAO aerosol at the most penetrating particle size (0.3 µm MMAD). Measure upstream and downstream concentrations with a calibrated photometer. Pass criteria: combined two-stage penetration less than 0.0001% (DF >= 1E6). Test at design airflow and after filter installation, replacement, or housing maintenance.
Rationale: In-situ DOP/PAO aerosol challenge at 0.3 µm MMAD validates HEPA filtration efficiency at the most penetrating particle size, confirming the ventilation system provides the required decontamination factor for airborne particulate discharges.
Test verification, ventilation, session-237
VER-METHODS-074 Verify SUB-REQS-009: After a representative decontamination cycle using the specified decontamination agents and procedures, perform surface contamination surveys on a minimum of 20 representative cell interior surfaces using direct-reading alpha and beta/gamma contamination monitors calibrated to relevant isotopes. Pass criteria: all surveyed surfaces below 4 Bq/cm2 alpha and 40 Bq/cm2 beta-gamma within 8 hours of decontamination start.
Rationale: Post-decontamination surface survey validates that cell internal surfaces can be cleaned to the required residual contamination level using standard procedures, confirming decontaminability for routine maintenance and end-of-life decommissioning.
Demonstration verification, hot-cell, session-237
VER-METHODS-075 Verify SUB-REQS-017: Perform hydrostatic pressure test on each active drain run at 1.5x design pressure. Confirm no leakage from inner pipe into the annular space over a 30-minute hold period using calibrated pressure gauges. Verify leak detection system functionality by injecting a known volume of tracer liquid into the annular monitoring sump. Pass criteria: zero leakage on pressure test; leak detection alarm activates within 60 seconds of tracer injection.
Rationale: Hydrostatic pressure testing at 1.5x design pressure validates drain pipe integrity including the double-containment annular space, confirming active liquid spills cannot reach the building drainage or ground without detection.
Test verification, effluent, session-237
VER-METHODS-076 Verify SUB-REQS-025: With the duty extract fan operating at design flow, simulate fan failure by tripping the duty fan motor. Measure time from duty fan trip to standby fan reaching design flow using calibrated flow measurement. Record pressure cascade measurements throughout changeover. Pass criteria: standby fan reaches design flow within the specified changeover time; pressure cascade does not reverse at any monitored zone during changeover.
Rationale: Fan changeover timing test validates the standby fan automatic start capability, confirming ventilation containment is maintained during single-fan failure. Uninterrupted extract flow prevents loss of negative pressure cascade.
Test verification, ventilation, session-237
VER-METHODS-077 Verify SUB-REQS-027: With the ventilation system operating at design conditions and all laboratory areas in normal configuration, measure differential pressure at each zone boundary using calibrated differential pressure transmitters. Verify corridor (ambient reference), general labs (-15 Pa), active labs (-30 Pa), fume cupboards (-50 Pa), and hot cells (-75 Pa). Pass criteria: all measured differentials within ±5 Pa of design values, sustained for 30 minutes minimum. Repeat with one fume cupboard sash fully open to verify cascade stability under transient load.
Rationale: Differential pressure measurement across all laboratory zones validates the pressure cascade from clean to active areas, confirming the designed airflow direction prevents contamination migration to lower-hazard zones.
Test verification, ventilation, session-237
VER-METHODS-078 Verify SUB-REQS-034: Place calibrated gamma check sources (Cs-137) at each area monitor location and confirm reading within ±20% of expected dose rate. Verify alarm trip points by incrementally increasing source activity. Confirm signal transmission to centralised monitoring display with timestamp accuracy within 5 seconds. Pass criteria: all monitors respond within ±20% of calibration source value; alarm thresholds trigger within 10% of set points; centralised display updates within 5 seconds.
Rationale: Gamma check source and alarm testing validates each area monitor's measurement accuracy and alarm function, confirming the radiation protection monitoring network provides reliable data for worker protection and regulatory compliance.
Test verification, rad-protection, session-237
VER-METHODS-079 Verify SYS-REQS-014: Commission automatic fire detection system in all active areas. Test detection response time with simulated fire sources in hot cell, waste store, and ductwork locations. Verify suppression agent discharge does not breach containment or spread contamination beyond the fire zone. Pass criteria: detection within 30 seconds of simulated fire source, suppression activation within 60 seconds, no measurable increase in airborne contamination outside the affected zone.
Rationale: Active commissioning testing of fire detection and suppression is required to demonstrate performance in the specific geometry and environment of each active area. Simulated fire sources validate detection sensitivity in the presence of radiological hazards. Contamination monitoring during suppression discharge confirms the suppression strategy does not create a secondary radiological hazard, which is the key differentiator from conventional fire suppression verification.
Test verification, fire-protection, session-238, duplicate-of-VER-METHODS-080
VER-METHODS-080 Verify SYS-REQS-014: Commission automatic fire detection system in all active areas. Test detection response time with simulated fire sources in hot cell, waste store, and ductwork locations. Verify suppression agent discharge does not breach containment or spread contamination beyond the fire zone. Pass: detection within 30 seconds, suppression within 60 seconds, no contamination spread.
Rationale: DUPLICATE: This verification entry duplicates VER-METHODS-079. Retained for traceability but superseded. See VER-METHODS-079 for the authoritative fire suppression verification method.
Test verification, fire-protection, session-238
VER-METHODS-081 Verify SYS-REQS-015: Simulate mains power failure during normal laboratory operations. Confirm UPS changeover completes within 10 seconds. Verify all safety-critical monitors continue operating without data loss. Run UPS on battery load for 4 hours. Pass: changeover under 10 seconds, zero monitoring gaps, 4-hour battery endurance.
Rationale: UPS changeover testing during live operations validates the actual transfer time experienced by safety-critical instruments, including transient effects on sensitive detectors. The 4-hour endurance test under representative load confirms battery sizing calculations and identifies any cells with degraded capacity. Zero monitoring gap verification ensures compliance with environmental permit conditions for continuous discharge monitoring.
Test verification, emergency-power, session-238
VER-METHODS-082 Verify SYS-REQS-016: Perform seismic qualification analysis of biological shielding, hot cell containment, ventilation ductwork connections, and safety-critical instrument mountings using site-specific 10000-year return period response spectra. Pass: all containment barriers and safety-critical mountings demonstrate positive margin with factor of safety exceeding 1.5.
Rationale: Seismic qualification by analysis rather than test is standard practice for building-scale nuclear structures where shake-table testing is impractical. The 1.5 safety factor on positive margin accounts for material property uncertainties, construction tolerances, and modelling assumptions in finite element analysis. Site-specific response spectra must be used rather than generic envelopes because UK dockyard sites have specific soil-structure interaction characteristics that affect amplification.
Analysis verification, seismic, session-238

Traceability Matrix — Verification

RequirementVerified ByDescription
SYS-REQS-013 VER-METHODS-079
SYS-REQS-016 VER-METHODS-082
SYS-REQS-015 VER-METHODS-081
SYS-REQS-014 VER-METHODS-080
IFC-DEFS-048 VER-METHODS-066
IFC-DEFS-052 VER-METHODS-070
IFC-DEFS-051 VER-METHODS-069
IFC-DEFS-050 VER-METHODS-068
IFC-DEFS-049 VER-METHODS-067
IFC-DEFS-001 VER-METHODS-001
IFC-DEFS-002 VER-METHODS-002
IFC-DEFS-003 VER-METHODS-003
IFC-DEFS-004 VER-METHODS-004
IFC-DEFS-005 VER-METHODS-005
IFC-DEFS-007 VER-METHODS-009
IFC-DEFS-008 VER-METHODS-010
IFC-DEFS-009 VER-METHODS-011
IFC-DEFS-010 VER-METHODS-012
IFC-DEFS-011 VER-METHODS-013
IFC-DEFS-012 VER-METHODS-014
IFC-DEFS-013 VER-METHODS-016
IFC-DEFS-014 VER-METHODS-017
IFC-DEFS-015 VER-METHODS-018
IFC-DEFS-016 VER-METHODS-019
IFC-DEFS-017 VER-METHODS-020
IFC-DEFS-018 VER-METHODS-021
IFC-DEFS-019 VER-METHODS-022
IFC-DEFS-020 VER-METHODS-023
IFC-DEFS-021 VER-METHODS-024
IFC-DEFS-022 VER-METHODS-027
IFC-DEFS-023 VER-METHODS-028
IFC-DEFS-024 VER-METHODS-029
IFC-DEFS-025 VER-METHODS-030
IFC-DEFS-026 VER-METHODS-031
IFC-DEFS-027 VER-METHODS-039
IFC-DEFS-028 VER-METHODS-040
IFC-DEFS-029 VER-METHODS-041
IFC-DEFS-030 VER-METHODS-042
IFC-DEFS-031 VER-METHODS-043
IFC-DEFS-032 VER-METHODS-044
IFC-DEFS-033 VER-METHODS-048
IFC-DEFS-034 VER-METHODS-049
IFC-DEFS-035 VER-METHODS-050
IFC-DEFS-036 VER-METHODS-051
IFC-DEFS-037 VER-METHODS-052
IFC-DEFS-038 VER-METHODS-053
IFC-DEFS-039 VER-METHODS-054
IFC-DEFS-040 VER-METHODS-055
IFC-DEFS-041 VER-METHODS-057
IFC-DEFS-042 VER-METHODS-058
IFC-DEFS-043 VER-METHODS-059
IFC-DEFS-044 VER-METHODS-060
IFC-DEFS-045 VER-METHODS-063
IFC-DEFS-046 VER-METHODS-064
IFC-DEFS-047 VER-METHODS-065
SUB-REQS-034 VER-METHODS-078
SUB-REQS-027 VER-METHODS-077
SUB-REQS-025 VER-METHODS-076
SUB-REQS-017 VER-METHODS-075
SUB-REQS-009 VER-METHODS-074
SUB-REQS-005 VER-METHODS-073
SUB-REQS-004 VER-METHODS-072
SUB-REQS-001 VER-METHODS-071
SUB-REQS-083 VER-METHODS-062
SUB-REQS-081 VER-METHODS-061
SUB-REQS-076 VER-METHODS-056
SUB-REQS-059 VER-METHODS-047
SUB-REQS-051 VER-METHODS-046
SUB-REQS-050 VER-METHODS-045
SUB-REQS-033 VER-METHODS-038
SUB-REQS-042 VER-METHODS-037
SUB-REQS-041 VER-METHODS-036
SUB-REQS-037 VER-METHODS-035
SUB-REQS-036 VER-METHODS-034
SUB-REQS-032 VER-METHODS-033
SUB-REQS-031 VER-METHODS-032
SUB-REQS-038 VER-METHODS-026
SUB-REQS-035 VER-METHODS-025
SUB-REQS-016 VER-METHODS-015
SUB-REQS-011 VER-METHODS-008
SUB-REQS-008 VER-METHODS-007
SUB-REQS-006 VER-METHODS-006