Surgical Robot Interim QC: Orphan Resolution and Haptic Cybersecurity Gap
System
{{entity:Surgical Robot System}} (se-surgical-robot), interim QC session triggered 4 sessions after LAST_QC_SESSION (351). Project at 213 requirements across 8 subsystems with decomposition in-progress on the {{entity:surgeon console}}. Focus: requirements created in sessions 353–354 (Surgeon Input Console ARC entries, Haptic Feedback Subsystem final requirements).
Findings
Orphan resolution (5 → 0). Lint identified 5 orphaned requirements. Three were architecture decisions from the past two sessions lacking trace links to the system requirements they support: {{stk:ARC-MAIN-009}} (Energy Delivery dual-modality), {{stk:ARC-MAIN-012}} (Surgeon Input Console E-stop separation), and {{stk:ARC-MAIN-014}} (Haptic galvanic isolation). {{stk:SUB-MAIN-065}} (IEC 60601-1 compliance for Haptic subsystem) was also unlinked.
Duplicate architecture decision. {{stk:ARC-MAIN-013}} was created 6 seconds after {{stk:ARC-MAIN-012}} in session 353 with truncated text but identical subject matter (Surgeon Input Console physical integration). ARC-MAIN-013 carried no trace links. Deleted.
Missing rationale and verification on ARC-MAIN-014. {{stk:ARC-MAIN-014}} (Haptic galvanic isolation architecture) was created in session 354 without --verification or --rationale fields. The text contained engineering content but the metadata fields were blank — a session-354 authoring error.
Cybersecurity gap on Console Computer. Lint finding 2 flagged digital components without cybersecurity requirements. The {{entity:console computer}} ({{hex:D0F51018}}) is the only subsystem with a hospital network interface (PACS/HIS integration) yet had no network isolation requirement. This is the only component that could bridge the hospital IT network to the CAN-FD surgical control bus.
Lint summary: 97 findings total (1 high, 32 medium, 64 low). Most medium findings are systemic: 8 regulated entities with no compliance requirements, 11 system-essential entities with no redundancy requirements, and 4 digital components without cybersecurity requirements. These are scope items for dedicated decomposition sessions, not addressable in a single QC pass.
Verification coverage: 48 VER entries against 102 SUB+IFC requirements (47%). The threshold for marking qc-reviewed is 50%; this interim QC session adds targeted coverage but the full 50% gate will be assessed at first-pass-complete.
Corrections
Linked 4 orphaned requirements:
- {{stk:ARC-MAIN-009}} → {{sys:SYS-MAIN-017}} (energy delivery architecture supports electrosurgical energy req)
- {{stk:ARC-MAIN-012}} → {{sys:SYS-MAIN-010}} (console E-stop architecture supports 50ms arrest requirement)
- {{stk:ARC-MAIN-014}} → {{sys:SYS-MAIN-004}} (galvanic isolation is the enabling architecture for force sensing)
- {{stk:SUB-MAIN-065}} → {{sys:SYS-MAIN-002}} (IEC 60601-1 certification is the regulatory path for Haptic safety integrity)
Fixed ARC-MAIN-014: Added verification: Inspection and rationale citing IEC 60601-1 leakage current limit (≤10µA CF-type) as the driver for the 4kVrms galvanic isolation barrier.
Deleted ARC-MAIN-013: Duplicate of ARC-MAIN-012. No trace links present on the duplicate; deletion logged here.
Added cybersecurity requirement REQ-SESURGICALROBOT-024: Console Computer SHALL implement network traffic isolation between the surgical control network and the hospital information network with no direct data path between safety-critical functions and external interfaces. Traced to {{sys:SYS-MAIN-002}}. Paired with penetration test VER entry REQ-SESURGICALROBOT-025.
Added STANDBY backdrive VER REQ-SESURGICALROBOT-028 for {{sub:SUB-MAIN-064}}: 10-position handle force sweep at ≤0.1N, plus mode-transition transient check (≤0.5N spike). Prior VER entries VER-MAIN-047 and VER-MAIN-048 (created session 354) covered {{sub:SUB-MAIN-062}} and {{sub:SUB-MAIN-063}} respectively; this session adds coverage for SUB-MAIN-064.
flowchart TB
FSM["Force Sensing Module"]
FSC["Force Signal Conditioner"]
HC["Haptic Controller"]
MHA["Master Handle Actuator"]
FSM -->|strain gauge signals| FSC
FSC -->|SPI 16-bit force data| HC
HC -->|CAN FD torque setpoints| MHA
Residual
Lint findings 2–13 (missing cybersecurity and compliance requirements on the remaining digital and regulated components) are not addressed. The {{entity:interlock subsystem}} ({{hex:40A51010}}), {{entity:motion control system}} ({{hex:51F73A18}}), and {{entity:energy delivery controller}} ({{hex:41B53B18}}) all lack compliance requirements against IEC 62061/62304. These require dedicated decomposition attention, not QC patching.
VER coverage at 47% (48/102 SUB+IFC) — below the 50% first-pass-complete gate. Four additional targeted VER entries would clear the gate; recommend addressing in the next Surgeon Input Console decomposition session.
Next
Decomposition continues with {{entity:surgeon console}} — the target set by session 353 remains undecomposed at the component level. Console Computer cybersecurity requirement establishes the first sub-component constraint. The next session should classify the Console Computer components (haptic master arms, stereo viewer, foot pedal assembly, console computer proper) in the SE:surgical-robot namespace, create SUB requirements for each, and push VER coverage above 50% to clear the first-pass-complete gate.